In Stentless bioprosthesis proximal joint of the prosthesis may loosen , causing blood between aortic wall and valve can come . When valve thrombosis there arises an increase in the gradient and an increase in the back flow ( leakage ) due to the larger mass, to enable the valve closes slowly. | In Stentless bioprosthesis proximal joint of the prosthesis may loosen , causing blood between aortic wall and valve can come . When valve thrombosis there arises an increase in the gradient and an increase in the back flow (leakage) due to the larger mass, to enable the valve closes slowly. | ||
==References== | ==References== |
Revision as of 13:04, 29 December 2013
Biological prosthetic valves
Valve | Type | Size (mm) | Peak gradient (mmHg) | Mean gradient (mmHg) | EOA (cm2) |
---|---|---|---|---|---|
Baxter Perimount | Stented bovine pericardial | 19 | 32.5 ± 8.5 | 19.5 ± 5.5 | 1.3 ± 0.2 |
21 | 24.9 ± 7.7 | 13.8 ± 4.0 | 1.3 ± 0.3 | ||
23 | 19.9 ± 7.4 | 11.5 ± 3.9 | 1.6 ± 0.3 | ||
25 | 16.5 ± 7.8 | 10.7 ± 3.8 | 1.6 ± 0.4 | ||
27 | 12.8 ± 5.4 | 4.8 ± 2.2 | 2.0 ± 0.4 | ||
Biocor | Stented porcine | 23 | 30.0 ± 10.7 | 20 ± 6.6 | 1.3 ± 0.3 |
25 | 23.0 ± 7.9 | 16 ± 5.1 | 1.7 ± 0.4 | ||
27 | 22.0 ± 6.5 | 15.0 ± 3.7 | 2.2 ± 0.4 | ||
BioFlo | Stented bovine pericardial | 19 | 37.2 ± 8.8 | 26.4 ± 5.5 | 0.7 ± 0.1 |
21 | 28.7 ± 6.2 | 18.7 ± 5.5 | 1.1 ± 0.1 | ||
Carpentier Edwards Pericardial | Stented bovine pericardial | 19 | 32.1 ± 3.4 | 24.2 ± 8.6 | 1.2 ± 0.3 |
21 | 25.7 ± 9.9 | 20.3 ± 9.1 | 1.5 ± 0.4 | ||
23 | 21.7 ± 8.6 | 13.0 ± 5.3 | 1.8 ± 0.3 | ||
25 | 16.5 ± 5.4 | 9.0 ± 2.3 | |||
Carpentier Edwards Standard | Stented porcine | 19 | 43.5 ± 12.7 | 25.6 ± 8.0 | 0.9 ± 0.2 |
21 | 27.7 ± 7.6 | 17.3 ± 6.2 | 1.5 ± 0.3 | ||
23 | 28.9 ± 7.5 | 16.1 ± 6.2 | 1.7 ± 0.5 | ||
25 | 24.0 ± 7.1 | 12.9 ± 4.6 | 1.9 ± 0.5 | ||
27 | 22.1 ± 8.2 | 12.1 ± 5.5 | 2.3 ± 0.6 | ||
29 | 9.9 ± 2.9 | 2.8 ± 0.5 | |||
Carpentier Supra - Annular | Stented porcine | 19 | 34.1 ± 2.7 | 1.1 ± 0.1 | |
21 | 28.0 ± 10.5 | 17.5 ± 3.8 | 1.4 ± 0.9 | ||
23 | 25.3 ± 10.5 | 13.4 ± 4.5 | 1.6 ± 0.6 | ||
25 | 24.4 ± 7.6 | 13.2 ± 4.8 | 1.8 ± 0.4 | ||
27 | 16.7 ± 4.7 | 8.8 ± 2.8 | 1.9 ± 0.7 | ||
CryoLife | Stentless | 19 | 9.0 ± 2.0 | 1.5 ± 0.3 | |
21 | 6.6 ± 2.9 | 1.7 ± 0.4 | |||
23 | 6.0 ± 2.3 | 2.3 ± 0.2 | |||
25 | 6.1 ± 2.6 | 2.6 ± 0.2 | |||
27 | 4.0 ± 2.4 | 2.8 ± 0.3 | |||
Extended Biocor | Stentless | 19 | 17.5 ± 6.5 | 9.6 ± 3.6 | 1.4 ± 0.4 |
21 | 17.5 ± 6.5 | 9.6 ± 3.6 | 1.4 ± 0.4 | ||
23 | 14.7 ± 7.3 | 7.7 ± 3.8 | 1.7 ± 0.4 | ||
25 | 14.0 ± 4.3 | 7.4 ± 2.5 | 1.8 ± 0.4 | ||
Hancock | Stented porcine | 21 | 18.0 ± 6.0 | 12.0 ± 2.0 | |
23 | 16.0 ± 2.0 | 11.0 ± 2.0 | |||
25 | 15.0 ± 3.0 | 10.0 ± 3.0 | |||
Hancock II | Stented porcine | 21 | 14.8 ± 4.1 | 1.3 ± 0.4 | |
23 | 34.0 ± 13.0 | 16.6 ± 8.5 | 1.3 ± 0.4 | ||
25 | 22.0 ± 5.3 | 10.8 ± 2.8 | 1.6 ± 0.4 | ||
29 | 16.2 ± 1.5 | 8.2 ± 1.7 | 1.6 ± 0.2 | ||
Homograft | Homograft valves | ? | ? | ? | ? |
Intact | Stented porcine | 19 | 40.4 ± 15.4 | 24.5 ± 9.3 | |
21 | 40.9 ± 15.6 | 19.6 ± 8.1 | 1.6 ± 0.4 | ||
23 | 32.7 ± 9.6 | 19.0 ± 6.1 | 1.6 ± 0.4 | ||
25 | 29.7 ± 15.0 | 17.7 ± 7.9 | 1.7 ± 0.3 | ||
27 | 25.0 ± 7.6 | 15.0 ± 4.5 | |||
Ionescu - Shiley | Stented bovine pericardial | 17 | 23.8 ± 3.4 | 0.9 ± 0.1 | |
19 | 19.7 ± 5.9 | 13.3 ± 3.9 | 1.1 ± 0.1 | ||
21 | 26.6 ± 9.0 | ||||
23 | 15.6 ± 4.4 | ||||
Labcor Santiago | Stented bovine pericardial | 19 | 18.6 ± 5.0 | 11.8 ± 3.3 | 1.2 ± 0.1 |
21 | 17.5 ± 6.6 | 8.2 ± 4.5 | 1.3 ± 0.1 | ||
23 | 14.8 ± 5.2 | 7.8 ± 2.9 | 1.8 ± 0.2 | ||
25 | 12.3 ± 3.4 | 6.8 ± 2.0 | 2.1 ± 0.3 | ||
Labcor Synergy | Stented porcine | 21 | 24.3 ± 8.1 | 13.3 ± 4.2 | 1.1 ± 0.3 |
23 | 27.3 ± 13.7 | 15.3 ± 6.9 | 1.4 ± 0.4 | ||
25 | 22.5 ± 11.9 | 13.2 ± 6.4 | 1.5 ± 0.4 | ||
27 | 17.8 ± 7.0 | 10.6 ± 4.6 | 1.8 ± 0.5 | ||
Medtronic Freestyle | Stentless | 19 | 13.0 ± 3.9 | ||
21 | 9.1 ± 5.1 | 1.4 ± 0.3 | |||
23 | 11.0 ± 4.0 | 8.1 ± 4.6 | 1.7 ± 0.5 | ||
25 | 5.3 ± 3.1 | 2.1 ± 0.5 | |||
27 | 4.6 ± 3.1 | 2.5 ± 0.1 | |||
Medtronic Mosaic | Stented porcine | 21 | 14.2 ± 5.0 | 1.4 ± 0.4 | |
23 | 23.8 ± 11.0 | 13.7 ± 4.8 | 1.5 ± 0.4 | ||
25 | 22.5 ± 10.0 | 11.7 ± 5.1 | 1.8 ± 0.5 | ||
27 | 10.4 ± 4.3 | 1.9 ± 0.1 | |||
29 | 11.1 ± 4.3 | 2.1 ± 0.2 | |||
Mitroflow | Stented bovine pericardial | 19 | 18.6 ± 5.3 | 13.1 ± 3.3 | 1.1 ± 0.2 |
Prima | Stentless | 21 | 28.8 ± 6.0 | 13.7 ± 1.9 | 1.4 ± 0.7 |
23 | 21.5 ± 7.5 | 11.5 ± 4.9 | 1.5 ± 0.3 | ||
25 | 22.1 ± 12.5 | 11.6 ± 7.2 | 1.8 ± 0.5 | ||
Sorin Pericarbon | Stentless | 19 | 36.5 ± 9.0 | 28.9 ± 7.3 | 1.2 ± 0.5 |
21 | 28.0 ± 13.3 | 23.8 ± 11.1 | 1.3 ± 0.6 | ||
23 | 27.5 ± 11.5 | 23.2 ± 7.6 | 1.5 ± 0.5 | ||
St. Jude Medical | Stentless | 21 | 22.6 ± 14.5 | 10.7 ± 7.2 | 1.3 ± 0.6 |
23 | 16.2 ± 9.0 | 8.2 ± 4.7 | 1.6 ± 0.6 | ||
25 | 12.7 ± 8.2 | 6.3 ± 4.1 | 1.8 ± 0.5 | ||
27 | 10.1 ± 5.8 | 5.0 ± 2.9 | 2.0 ± 0.3 | ||
29 | 7.7 ± 4.4 | 4.1 ± 2.4 | 2.4 ± 0.6 |
Valve | Type | Size (mm) | Peak gradient (mmHg) | Mean gradient (mmHg) | Peak Velocity (m/s) | Pressure half-time (ms) | EOA (cm2) |
---|---|---|---|---|---|---|---|
Biocor | Stentless bioprosthesis 27 13 ± 1 | ||||||
29 14 ± 2.5 | |||||||
31 11.5 ± 0.5 | |||||||
33 12 ± 0.5 | |||||||
Stented pericardial bioprosthesis BioFlo 25 10 ± 2 6.3 ± 1.5 0.1 ± 2 | |||||||
27 9.5 ± 2.6 5.4 ± 1.2 2 ± 0.3 | |||||||
29 5 ± 2.8 3.6 ± 0.2 2.4 ± 1 | |||||||
31 4.0 2.0 2.3 | |||||||
Carpentier - Edwards stented bioprosthesis 27 6 ± 2 1.7 ± 0.3 98 ± 28 | |||||||
29 4.7 2 ± 1.76 ± 92 ± 14 12:27 | |||||||
4.4 31 ± 2 92 ± 1:54 ± 0:15 19 | |||||||
6 33 ± 3 93 ± 12 | |||||||
Carpentier - Edwards pericardial stented Bioprosthesis 27 3.6 1.6 100 | |||||||
29 5:25 2:36 ± 1.67 ± 0.3 110 ± 15 | |||||||
31 4:05 1:53 ± 0.83 ± 0.1 90 ± 11 | |||||||
23 1.0 0.8 80 | |||||||
Hancock I ( or NOS ) stented bioprosthesis 27 10 ± 4 5 ± 2 1.3 ± 0.8 | |||||||
29 7 ± 3 2:46 ± 0.79 115 ± 20 1.5 ± 0.2 | |||||||
31 4 ± 0.86 4.86 ± 1.69 95 ± 17 1.6 ± 0.2 | |||||||
33 3 2 ± 3.87 ± 2 90 ± 12 1.9 ± 0.2 | |||||||
Hancock II stented bioprosthesis 27 2:21 ± 0:14 | |||||||
29 2.77 ± 0:11 | |||||||
31 2.84 ± 0.1 | |||||||
33 3:15 ± 0:22 | |||||||
Hancock pericardial stented bioprosthesis 29 2.61 ± 1:39 1:42 ± 0:14 105 ± 36 | |||||||
31 3:57 1:02 1:51 ± 0:27 ± 81 ± 23 | |||||||
Ionescu - Shiley stented bioprosthesis 25 4.87 ± 1.43 ± 1:08 ± 0:15 93 11 | |||||||
27 3:21 1:31 ± 0:24 ± 0.82 100 ± 28 | |||||||
29 3:22 0:57 ± 1:38 ± 0.2 85 ± 8 | |||||||
31 ± 0.9 1.45 ± 3.63 100 ± 36 00:06 | |||||||
Ionescu - Shiley low profile stented bioprosthesis 29 3:31 1:36 ± 0.96 ± 0.25 80 ± 30 | |||||||
31 2.74 ± 0:37 1:33 ± 0:14 ± 79 15 | |||||||
Stented pericardial bioprosthesis Labcor - Santiago 25 8.7 4.5 97 2.2 | |||||||
27 5.6 ± 2.3 2.8 ± 1.5 85 ± 18 ± 2:12 0:48 | |||||||
29 ± 6.2 2.1 ± 1.3 80 ± 3 34 ± 0.73 2:11 | |||||||
Medtronic Intact Porcine stented bioprosthesis 29 3.5 ± 1.6 ± 0:51 0:22 | |||||||
31 ± 4.2 1.6 ± 1:44 0:26 | |||||||
33 4 ± 1.3 1.4 ± 0:24 | |||||||
35 3.2 ± 1.77 1.3 ± 0.5 | |||||||
Stented bioprosthesis Mitroflow 25 6.9 2.0 90 | |||||||
27 3:07 0.91 ± 1.5 90 ± 20 | |||||||
29 ± 3.5 1.43 ± 1.65 102 ± 21 12:29 | |||||||
31 3.85 ± 0.81 91 ± 1:32 ± 0:26 22 | |||||||
Stentless quadrileaflet bovine pericardial bioprosthesis Stentless 26 2.2 1.7 ± 1.6 103 ± 31 1.7 | |||||||
28 1:58 ± 0.25 1.7 ± 0.6 | |||||||
30 1:42 0:32 ± 2.3 ± 0.4 | |||||||
Wessex stented bioprosthesis 29 3.69 ± 0.61 1.66 ± 83 ± 19 00:17 | |||||||
31 3:31 1:41 ± 0.83 ± 0.25 80 ± 21 |
Related valve dysfunction in bioprostheses
In Stentless bioprosthesis proximal joint of the prosthesis may loosen , causing blood between aortic wall and valve can come . When valve thrombosis there arises an increase in the gradient and an increase in the back flow (leakage) due to the larger mass, to enable the valve closes slowly.
References
- Zoghbi WA, Chambers JB, Dumesnil JG, Foster E, Gottdiener JS, Grayburn PA, Khandheria BK, Levine RA, Marx GR, Miller FA Jr, Nakatani S, Quiñones MA, Rakowski H, Rodriguez LL, Swaminathan M, Waggoner AD, Weissman NJ, Zabalgoitia M, American Society of Echocardiography's Guidelines and Standards Committee, Task Force on Prosthetic Valves, American College of Cardiology Cardiovascular Imaging Committee, Cardiac Imaging Committee of the American Heart Association, European Association of Echocardiography, European Society of Cardiology, Japanese Society of Echocardiography, Canadian Society of Echocardiography, American College of Cardiology Foundation, American Heart Association, European Association of Echocardiography, European Society of Cardiology, Japanese Society of Echocardiography, and Canadian Society of Echocardiography. Recommendations for evaluation of prosthetic valves with echocardiography and doppler ultrasound: a report From the American Society of Echocardiography's Guidelines and Standards Committee and the Task Force on Prosthetic Valves, developed in conjunction with the American College of Cardiology Cardiovascular Imaging Committee, Cardiac Imaging Committee of the American Heart Association, the European Association of Echocardiography, a registered branch of the European Society of Cardiology, the Japanese Society of Echocardiography and the Canadian Society of Echocardiography, endorsed by the American College of Cardiology Foundation, American Heart Association, European Association of Echocardiography, a registered branch of the European Society of Cardiology, the Japanese Society of Echocardiography, and Canadian Society of Echocardiography. J Am Soc Echocardiogr. 2009 Sep;22(9):975-1014; quiz 1082-4. DOI:10.1016/j.echo.2009.07.013 |