Mechanical
Mechanical prosthetic valves
Valve | Type | Size (mm) | Peak gradient (mmHg) | Mean gradient (mmHg) | EOA (cm2) |
---|---|---|---|---|---|
ATS | Bileaflet | 19 | 47.0 ± 12.6 | 25.3 ± 8 | 1.1 ± 0.3 |
21 | 23.7 ± 6.8 | 15.9 ± 5.0 | 1.4 ± 0.5 | ||
23 | 14.4 ± 4.9 | 1.7 ± 0.5 | |||
25 | 11.3 ± 3.7 | 2.1 ± 0.7 | |||
27 | 8.4 ± 3.7 | 2.5 ± 0.1 | |||
29 | 8.0 ± 3.0 | 3.1 ± 0.8 | |||
18 | 21.0 ± 1.8 | 1.2 ± 0.3 | |||
20 | 21.4 ± 4.2 | 11.1 ± 3.5 | 1.3 ± 0.3 | ||
22 | 18.7 ± 8.3 | 10.5 ± 4.5 | 1.7 ± 0.4 | ||
24 | 15.1 ± 5.6 | 7.5 ± 3.1 | 2.0 ± 0.6 | ||
26 | 6.0 ± 2.0 | 2.1 ± 0.4 | |||
Bjork-Shiley | Single tilting disc | 21 | 38.9 ± 11.9 | 21.8 ± 3.4 | 1.1 ± 0.3 |
23 | 28.8 ± 11.2 | 15.7 ± 5.3 | 1.3 ± 0.3 | ||
25 | 23.7 ± 8.2 | 13.0 ± 5.0 | 1.5 ± 0.4 | ||
27 | 10.0 ± 2.0 | 1.6 ± 0.3 | |||
Carbomedics Reduced | Bileaflet | 19 | 43.4 ± 1.2 | 24.4 ± 1.2 | 1.2 ± 0.1 |
Carbomedics Standard | Bileaflet | 19 | 38.0 ± 12.8 | 18.9 ± 8.3 | 1.0 ± 0.3 |
21 | 26.8 ± 10.1 | 12.9 ± 5.4 | 1.5 ± 0.4 | ||
23 | 22.5 ± 7.4 | 11.0 ± 4.6 | 1.4 ± 0.3 | ||
25 | 19.6 ± 7.8 | 9.1 ± 3.5 | 1.8 ± 0.4 | ||
27 | 17.5 ± 7.1 | 7.9 ± 3.2 | 2.2 ± 0.2 | ||
29 | 9.1 ± 4.7 | 5.6 ± 3.0 | 3.2 ± 1.6 | ||
Carbomedics Tophat | Bileaflet | 21 | 30.2 ± 10.9 | 14.9 ± 5.4 | 1.2 ± 0.3 |
23 | 24.2 ± 7.6 | 12.5 ± 4.4 | 1.4 ± 0.4 | ||
25 | 9.5 ± 2.9 | 1.6 ± 0.32 | |||
Edwards Duromedics | Bileaflet | 21 | 39.0 ± 13 | ||
23 | 32.0 ± 8.0 | ||||
25 | 26.0 ± 10.0 | ||||
27 | 24.0 ± 10.0 | ||||
Edwards Mira | Bileaflet | 19 | 18.2 ± 5.3 | 1.2 ± 0.4 | |
21 | 13.3 ± 4.3 | 1.6 ± 0.4 | |||
23 | 14.7 ± 2.8 | 1.6 ± 0.6 | |||
25 | 13.1 ± 3.8 | 1.9 | |||
MCRI On-X | Bileaflet | 19 | 21.3 ± 10.8 | 11.8 ± 3.4 | 1.5 ± 0.2 |
21 | 16.4 ± 5.9 | 9.9 ± 3.6 | 1.7 ± 0.4 | ||
23 | 15.9 ± 6.4 | 8.6 ± 3.4 | 1.9 ± 0.6 | ||
25 | 16.5 ± 10.2 | 6.9 ± 4.3 | 2.4 ± 0.6 | ||
Medtronic Advantage | Bileaflet | 23 | 10.4 ± 3.1 | 2.2 ± 0.3 | |
25 | 9.0 ± 3.7 | 2.8 ± 0.6 | |||
27 | 7.6 ± 3.6 | 3.3 ± 0.7 | |||
29 | 6.1 ± 3.8 | 3.9 ±0.7 | |||
Medtronic Hall | Single tilting disc | 20 | 34.4 ± 13.1 | 17.1 ± 5.3 | 1.2 ± 0.5 |
21 | 26.9 ± 10.5 | 14.1 ± 5.9 | 1.1 ± 0.2 | ||
23 | 26.9 ± 8.9 | 13.5 ± 4.8 | 1.4 ± 0.4 | ||
25 | 17.1 ± 7.0 | 9.5 ± 4.3 | 1.5 ± 0.5 | ||
27 | 18.9 ± 9.7 | 8.7 ± 5.6 | 1.9 ± 0.2 | ||
Monostrut Bjork-Shiley | Single tilting disc | 19 | 27.4 ± 8.8 | ||
21 | 27.5 ± 3.1 | 20.5 ± 6.2 | |||
23 | 20.3 ± 0.7 | 17.4 ± 6.4 | |||
25 | 16.1 ± 4.9 | ||||
27 | 11.4 ± 3.8 | ||||
Omnicarbon | Single tilting disc | 21 | 37.4 ± 12.8 | 20.4 ± 5.4 | 1.3 ± 0.5 |
23 | 28.8 ± 9.1 | 17.4 ± 4.9 | 1.5 ± 0.3 | ||
25 | 23.7 ± 8.1 | 13.2 ± 4.6 | 1.9 ± 0.5 | ||
27 | 20.1 ± 4.2 | 12.4 ± 2.9 | 2.1 ± 0.4 | ||
Omniscience | Single tilting disc | 21 | 50.8 ± 2.8 | 28.2 ± 2.2 | 0.9 ± 0.1 |
23 | 39.8 ± 8.7 | 20.1 ± 5.1 | 1.0 ± 0.1 | ||
Sorin Bicarbon | Bileaflet | 19 | 30.1 ± 4.5 | 16.7 ± 2.0 | 1.4 ± 0.1 |
21 | 22.0 ± 7.1 | 10.0 ± 3.3 | 1.2 ± 0.4 | ||
23 | 16.8 ± 6.1 | 7.7 ± 3.3 | 1.5 ± 0.2 | ||
25 | 11.2 ± 3.1 | 5.6 ± 1.6 | 2.4 ± 0.3 | ||
St. Jude Medical Haem Plus | Bileaflet | 19 | 28.5 ± 10.7 | 17.0 ± 7.8 | 1.9 ± 0.1 |
21 | 16.3 ± 17.0 | 10.6 ± 5.1 | 1.8 ± 0.5 | ||
23 | 16.8 ± 7.3 | 12.1 ± 4.2 | 1.7 ± 0.5 | ||
St. Jude Medical Regent | Bileaflet | 19 | 20.6 ± 12 | 11.0 ± 4.9 | 1.6 ± 0.4 |
21 | 15.6 ± 9.4 | 8.0 ± 4.8 | 2.0 ± 0.7 | ||
23 | 12.8 ± 6.8 | 6.9 ± 3.5 | 2.3 ± 0.9 | ||
25 | 11.7 ± 6.8 | 5.6 ± 3.2 | 2.5 ± 0.8 | ||
27 | 7.9 ± 5.5 | 3.5 ± 1.7 | 3.6 ± 0.5 | ||
St. Jude Medical Standard | Bileaflet | 19 | 42.0 ± 10.0 | 24.5 ± 5.8 | 1.5 ± 0.1 |
21 | 25.7 ± 9.5 | 15.2 ± 5.0 | 1.4 ± 0.4 | ||
23 | 21.8 ± 7.5 | 13.4 ± 5.6 | 1.6 ± 0.4 | ||
25 | 18.9 ± 7.3 | 11.0 ± 5.3 | 1.9 ± 0.5 | ||
27 | 13.7 ± 4.2 | 8.4 ± 3.4 | 2.5 ± 0.4 | ||
29 | 13.5 ± 5.8 | 7.0 ± 1.7 | 2.8 ± 0.5 | ||
Starr Edwards | Caged ball | 23 | 32.6 ± 12.8 | 22.0 ± 9.0 | 1.1 ± 0.2 |
24 | 34.1 ± 10.3 | 22.1 ± 7.5 | 1.1 ± 0.3 | ||
26 | 31.8 ± 9.0 | 19.7 ± 6.1 | |||
27 | 30.8 ± 6.3 | 18.5 ± 3.7 | |||
29 | 29.0 ± 9.3 | 16.3 ± 5.5 |
Valve | Type | Size (mm) | Peak gradient (mmHg) | Mean gradient (mmHg) | Peak Velocity (m/s) | Pressure half-time (ms) | EOA (cm2) |
---|---|---|---|---|---|---|---|
Bjork-Shiley | Tilting disc | 23 | 1.7 | 115 | |||
25 | 12 ± 4 | 6 ± 2 | 1.75 ± 0.38 | 99 ± 27 | 1.72 ± 0.6 | ||
27 | 10 ± 4 | 5 ± 2 | 1.6 ± 0.49 | 89 ± 28 | 1.81 ± 0.54 | ||
29 | 7.83 ± 2.93 | 2.83 ± 1.27 | 1.37 ± 0.25 | 79 ± 17 | 2.1 ± 0.43 | ||
31 | 6 ± 3 | 2 ± 1.9 | 1.41 ± 0.26 | 70 ± 14 | 2.2 ± 0.3 | ||
Bjork-Shiley monostrut | Tilting disc | 23 | 5.0 | 1.9 | |||
25 | 13 ± 2.5 | 5.57 ± 2.3 | 1.8 ± 0.3 | ||||
27 | 12 ± 2.5 | 4.53 ± 2.2 | 1.7 ± 0.4 | ||||
29 | 13 ± 3 | 4.26 ± 1.6 | 1.6 ± 0.3 | ||||
31 | 14 ± 4.5 | 4.9 ± 1.6 | 1.7 ± 0.3 | ||||
Carbomedics | Bileaflet | 23 | 1.9 ± 0.1 | 126 ± 7 | |||
25 | 10.3 ± 2.3 | 3.6 ± 0.6 | 1.3 ± 0.1 | 93 ± 8 | 2.9 ± 0.8 | ||
27 | 8.79 ± 3.46 | 3.46 ± 1.03 | 1.61 ± 0.3 | 89 ± 20 | 2.9 ± 0.75 | ||
29 | 8.78 ± 2.9 | 3.39 ± 0.97 | 1.52 ± 0.3 | 88 ± 17 | 2.3 ± 0.4 | ||
31 | 8.87 ± 2.34 | 3.32 ± 0.87 | 1.61 ± 0.29 | 92 ± 24 | 2.8 ± 1.14 | ||
33 | 8.8 ± 2.2 | 4.8 ± 2.5 | 1.5 ± 0.2 | 93 ± 12 | |||
Duromedics | Bileaflet | 27 | 13 ± 6 | 5 ± 3 | 1.61 ± 0.4 | 75 ± 12 | |
29 | 10 ± 4 | 3 ± 1 | 1.40 ± 0.25 | 85 ± 22 | |||
31 | 10.5 ± 4.33 | 3.3 ± 1.36 | 1.38 ± 0.27 | 81 ± 12 | |||
33 | 11.2 | 2.5 | 85 | ||||
Lillehei-Kaster | Tilting disc | 18 | 1.7 | 140 | |||
20 | 1.7 | 67 | |||||
22 | 1.56 ± 0.09 | 94 ± 22 | |||||
25 | 1.38 ± 0.27 | 124 ± 46 | |||||
Medtronic Hall | Tilting disc | 27 | 1.4 | 78 | |||
29 | 1.57 ± 0.1 | 69 ± 15 | |||||
31 | 1.45 ± 0.12 | 77 ± 17 | |||||
Omnicarbon | Tilting disc | 23 | 8.0 | ||||
25 | 6.05 ± 1.81 | 1.77 ± 0.24 | 102 ± 16 | ||||
27 | 4.89 ± 2.05 | 1.63 ± 0.36 | 105 ± 33 | ||||
29 | 4.93 ± 2.16 | 1.56 ± 0.27 | 120 ± 40 | ||||
31 | 4.18 ± 1.4 | 1.3 ± 0.23 | 134 ± 31 | ||||
33 | 4 ± 2 | ||||||
On-X | Bileaflet | 25 | 11.5 ± 3.2 | 5.3 ± 2.1 | 1.9 ± 1.1 | ||
27 | 10.3 ± 4.5 | 4.5 ± 1.6 | 2.2 ± 0.5 | ||||
29 | 10.3 ± 4.5 | 4.5 ± 1.6 | 2.2 ± 0.5 | ||||
31 | 9.8 ± 3.8 | 4.8 ± 2.4 | 2.5 ± 1.1 | ||||
33 | 9.8 ± 3.8 | 4.8 ± 2.4 | 2.5 ± 1.1 | ||||
Sorin Allcarbon | Tilting disc | 25 | 15 ± 0.25 | 4 ± 0.5 | 1.95 ± 0.02 | 70 ± 1 | |
27 | 11 ± 2.75 | 4 ± 0.5 | 1.65 ± 0.21 | 82 ± 20 | |||
29 | 12 ± 3 | 4 ± 1.25 | 1.73 ± 0.22 | 80 ± 14 | |||
31 | 10 ± 1.5 | 4 ± 1 | 1.66 ± 0.11 | 83 ± 14 | |||
St. Jude Medical | Bileaflet | 23 | 4.0 | 1.5 | 160 | 1.0 | |
25 | 2.5 ± 1 | 1.34 ± 1.12 | 75 ± 4 | 1.35 ± 0.17 | |||
27 | 11 ± 4 | 5 ± 1.82 | 1.61 ± 0.29 | 75 ± 10 | 1.67 ± 0.17 | ||
29 | 10 ± 3 | 4.15 ± 1.8 | 1.57 ± 0.29 | 85 ± 10 | 1.75 ± 0.24 | ||
31 | 12 ± 6 | 4.46 ± 2.22 | 1.59 ± 0.33 | 74 ± 13 | 2.03 ± 0.32 | ||
Starr-Edwards | Caged ball | 26 | 10.0 | 1.4 | |||
28 | 7 ± 2.75 | 1.9 ± 0.57 | |||||
30 | 12.2 ± 4.6 | 6.99 ± 2.5 | 1.7 ± 0.3 | 125 ± 25 | 1.65 ± 0.4 | ||
32 | 11.5 ± 4.2 | 5.08 ± 2.5 | 1.7 ± 0.3 | 110 ± 25 | 1.98 ± 0.4 | ||
34 | 5.0 | 2.6 |
Leakage
There is slight insufficiency, in all of the mechanical prosthetic valves it is the flow that is needed to allow the valve can be closed and keep the valve free from clots. This flow is due to the hinges of the valve. The degree of leakage and lekpatroon is by design prosthetic valve depends. At a St Jude can sometimes five small flames are observed. TEE This leakage is generally low.
Mitral regurgitation with prosthetic valve
Hemodynamically significant (at least moderate) mitral regurgitation with prosthetic valves without obstruction. ( PHT <120 msec)
Peak inflow velocity > 1.9 m/s (CW - Doppler)
Peakgradiënt > 14 mmHg (CW - Doppler)
TVI MV inflow (CW - Doppler/TVI LVOT (PW Doppler) > 2.5
These measurements can help you on the spot. Alert pathological mitraliskunstklep leakage after further research is needed with TEE.
Related valve dysfunction with mechanical prostheses
Bjork- Shiley: strut fracture
Carbomedics: degeneration disc
Hancock bioprosthesis: Distortion stents causing stenosis
Carpentier Edwards bioprosthesis: tear/holes in leaflets and development of stenosis
References
- Error fetching PMID 19733789: